DS5
- Quick and Easy HbA1c measurement
|
||||
|
|
HbA1c measurement in only 10 minutes |
|
|
Time to first result is only 7 minutes |
|
|
Immediate labile fraction removal using the DS5 Dispenser |
|
|
Simple to use system with icon driven touchscreen |
|
|
15 position autosampler |
|
|
Detects presence of HbS or HbC |
|
|
|
|
|
Intended Use The DS5 instrument and associated reagent kit is intended for the in vitro measurement of HbA1c in blood samples. The system's small size and ease of use make it ideal for main laboratory, clinic or satellite laboratory settings. Scientific principle The DS5 uses cation exchange chromatography in conjunction with gradient elution to separate human haemoglobin subtypes and variants from haemolyzed whole blood. The separated haemoglobin fractions are monitored by means of absorption of light and the chromatogram obtained is recorded and stored by the internal computer. A software program performs the analysis of the chromatogram and generates a results report on the integral thermal printer. Performance Data |
| Linearity | |
| The DS5 produces linear results between 4.15 and 14.9% HbA1c. | |
|
Within
batch precision
|
HbA1c | HbA1c |
|
Number
|
10 | 10 |
|
Mean
value
|
5.0 | 13.2 |
|
SD
|
0.11 | 0.26 |
|
CV
|
2.3% | 2.0% |
|
Between
batch precision
|
HbA1c | HbA1c |
|
Number
|
10 | 10 |
|
Mean
value
|
4.3 | 7.0 |
|
SD
|
0.08 | 0.27 |
|
CV
|
1.9% | 3.9% |
| Technical data | |
| Data Storage | |
| . | Storage for 15 chromatograms, results and summary. RS232 Unidirectional Serial Interface Port |
| Operation: | 15ºC to 35ºC, 10% - 80% non-condensing humidity | |
| Transportation: | -5ºC to 40ºC, 5% - 80% non-condensing humidity | |
| Voltage: | 110/115V or 220/240V; 50/60Hz | |
| Dimensions: | 27cm(w) x 41cm(d) x 35cm(h) | |
| Weight: | 20kg | |
| Ordering: | 274-000 DS5 Analyzer | |
| 704-000 DS5 Dispenser | ||
| 014-200 DS5 100 Test Kit | ||
|
Manufactured
by Drew Scientific
Ltd, United Kingdom
|